ASTM D2269-10(2020) - 1.11.2020
 
Significance and Use

5.1?The ultraviolet absorption of white mineral oils is used to determine their suitability for use in food, drug, and cosmetic applications.

5.2?The U.S. Pharmacopeia and the National Formulary specifications for mineral oil require the measurement of ultraviolet absorption.

 
1. Scope

1.1?This test method describes a procedure for the examination and evaluation of NF and USP grade white mineral oils.

1.2?This test method is not applicable to oils containing additives soluble in dimethyl sulfoxide (DMSO) that exhibit fluorescence or fluorescence quenching properties.

1.3?The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.4?This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. For specific warning statements, see 7.1.1 7.1.3.

1.5?This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

 
2. Referenced Documents

U.S. Pharmacopeia USP XIII/National Formulary

(NF XVIII) Available from The United States Pharmacopeia (USP), 12601 Twinbrook Parkway, Rockville, MD 20852.

D1840-22

Standard Test Method for Naphthalene Hydrocarbons in Aviation Turbine Fuels by Ultraviolet Spectrophotometry

E131-10(2023)

Standard Terminology Relating to Molecular Spectroscopy

E275-08(2022)

Standard Practice for Describing and Measuring Performance of Ultraviolet and Visible Spectrophotometers