ASTM D4785-08 - 1.4.2008
 
Significance and Use

This test method was developed for measuring low levels of iodine-131 in water. The results of the test may be used to determine if the concentration of iodine-131 in the sample exceeds the regulatory statutes for drinking water. With a suitable counting technique, sample size, and counting time, a detection limit of less than 0.037 Bq/L (1 pCi/L) is attainable by gamma-ray spectroscopy.

 
1. Scope

1.1 This test method covers the quantification of low levels of radioactive iodine in water by means of chemical separation and counting with a high-resolution gamma ray detector. Iodine is chemically separated from a 4-L water sample using ion exchange and solvent extraction and is then precipitated as cuprous iodide for counting.

1.2 The values stated in SI units are to be regarded as standard. The values given in parentheses are provided for information purposes only.

This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazard statements, see Note 2 , Note 3, Note 9, and Section 9 .

 
2. Referenced Documents

D3649-23

Standard Practice for High-Resolution Gamma-Ray Spectrometry of Water

D3648-23

Standard Practices for the Measurement of Radioactivity

D3370-25

Standard Practices for Sampling Water from Flowing Process Streams

D2777-21

Standard Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water

D1129-24

Standard Terminology Relating to Water

D1193-24

Standard Specification for Reagent Water

D5847-22

Standard Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis

D3856-11

Standard Guide for Management Systems in Laboratories Engaged in Analysis of Water