ASTM D5907-09 - 1.5.2009
 
Significance and Use

Solids, both as filterable and nonfilterable matter, are important in the treating of raw water and wastewater, and in monitoring of streams.

Waste solids impose a suspended and settleable residue in receiving waters. Suspended and soluble materials provide a matrix for some biological slime and, in sufficient quantity, impair respiration of organisms. These solids may create nuisance slime beds and odors while imposing a long-term biological oxidation load over limited receiving water areas.

Knowledge of suspended and soluble materials is important in treating raw water supplies. Knowledge of solids loading can aid in determining the type or amount of treatment, or both, necessary to make the water acceptable for use. Such information may also be used to determine acceptability of water after treatment. Too little treatment may not be desirable and excess treatment costs money.

Stream monitoring is important for environmental reasons. Stream improvements, water pollution monitoring, mass wasting, algal studies, and sediment loads are but a few of the many reasons streams are monitored.

 
1. Scope

1.1 This test method covers the determination of filterable and nonfilterable matter in drinking, surface, and saline waters, domestic and industrial wastes. The practical range of the determination of nonfilterable particulate matter is 4 to 20 000 mg/L. The practical range of the determination of filterable matter is 10 to 20 000 mg/L.

1.2 Since the results measured by this test are operationally defined, careful attention must be paid to following the procedure as specified.

1.3 This method for the determination of nonfilterable matter (TSS) must not be used when water samples were collected from open channel flow. For the determination of matter collected in open channel flow use Test Methods D 3977.

1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For a specific hazard statement, see Section 9.

 
2. Referenced Documents

D2777-21

Standard Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water

D3370-25

Standard Practices for Sampling Water from Flowing Process Streams

D3856-11

Standard Guide for Management Systems in Laboratories Engaged in Analysis of Water

D3977-97(2019)

Standard Test Methods for Determining Sediment Concentration in Water Samples

D596-01(2018)

Standard Guide for Reporting Results of Analysis of Water

D1129-24

Standard Terminology Relating to Water

D1192-98

Standard Guide for Equipment for Sampling Water and Steam in Closed Conduits (Withdrawn 2003)

D4411-03(2019)

Standard Guide for Sampling Fluvial Sediment in Motion

D5847-22

Standard Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis

D5905-98(2018)

Standard Practice for the Preparation of Substitute Wastewater

E1601-19

Standard Practice for Conducting an Interlaboratory Study to Evaluate the Performance of an Analytical Method

E319-85(1997)

Standard Practice for the Evaluation of Single-Pan Mechanical Balances

E898-20

Standard Practice for Calibration of Non-Automatic Weighing Instruments

D1193-24

Standard Specification for Reagent Water

D1429-13

Standard Test Methods for Specific Gravity of Water and Brine (Withdrawn 2022)