ASTM D5996-05 - 1.6.2005
 
1. Scope

1.1 This test method covers on-line analysis of high-purity water by the ion chromatography technique. This test method is applicable for measuring various anionic contaminants in high-purity water, typically in the range of 0.01 to 100 g/L. This test method is used to determine the concentration of acetate, formate, chloride, fluoride, phosphate, nitrate, and sulfate in a continuously flowing sample. The range of the test method is only as good as the reagent water available for preparing standards. At extremely low concentrations, 1.0 g/L, preparing standards is difficult, and extra care must be taken in their preparation. The sample may have to be conditioned from higher pressures and temperatures to conditions that are suitable for use by on-line instruments.

1.2 Online sample analysis of flowing streams does not lend itself to collaborative studies due to the nature of the sample and the possibility of contamination that may result from handling the sample as part of the collaborative study. Therefore this standard test method is not based on the results of a collaborative study but is intended to provide the best possible guidance for doing this type of analysis.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

D4453-17(2025)

Standard Practice for Handling of High Purity Water Samples

D3864-12(2021)

Standard Guide for On-Line Monitoring Systems for Water Analysis

D3370-25

Standard Practices for Sampling Water from Flowing Process Streams

D2777-21

Standard Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water

D1193-24

Standard Specification for Reagent Water

D1192-98

Standard Guide for Equipment for Sampling Water and Steam in Closed Conduits (Withdrawn 2003)

D1129-24

Standard Terminology Relating to Water

D1066-18(2025)

Standard Practice for Sampling Steam

D5847-22

Standard Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis

D5542-16

Standard Test Methods for Trace Anions in High Purity Water by Ion Chromatography (Withdrawn 2025)

D5810-96(2026)

Standard Guide for Spiking into Aqueous Samples