ASTM D6071-96(2000) - 1.1.2000
 
1. Scope

1.1 This test method covers the determination of trace sodium in high purity water. The method range of concentration is 1 to 40 µg/L sodium. The analyst may extend the range once its applicability has been ascertained.

Note 1-- It is necessary to perform replicate analysis and take an average to accurately determine values at the lower end of the stated range.

1.2 This method is a graphite furnace atomic absorption spectrophotometric method for the determination of trace sodium impurities in ultra high purity water.

1.3 this test method has been used successfully with a high purity water matrix. It is the responsibility of the analyst to determine the suitability of the test method for other matrices.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

D3919-15

Standard Practice for Measuring Trace Elements in Water by Graphite Furnace Atomic Absorption Spectrophotometry (Withdrawn 2024)

D3370-25

Standard Practices for Sampling Water from Flowing Process Streams

D2777-21

Standard Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water

D1129-24

Standard Terminology Relating to Water

D1193-24

Standard Specification for Reagent Water

D5847-22

Standard Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis

D5810-96(2026)

Standard Guide for Spiking into Aqueous Samples

D4453-17(2025)

Standard Practice for Handling of High Purity Water Samples

D1192-98

Standard Guide for Equipment for Sampling Water and Steam in Closed Conduits (Withdrawn 2003)

D1066-18(2025)

Standard Practice for Sampling Steam