ASTM D6889-03 - 10.3.2003
 
1. Scope

1.1 This practice covers a procedure for the screening of trace levels of volatile organic compounds in water samples by headspace solid phase microextraction (SPME) in combination with fast gas chromatography with flame ionization detection.

1.2 The results from this screening procedure are used to estimate analyte concentrations to prevent contamination of purge and trap or headspace analytical systems.

1.3 The compounds of interest must have a greater affinity for the SPME absorbent polymer or adsorbent than the sample matrix or headspace phase in which they reside.

1.4 Not all of the analytes which can be determined by SPME are addressed in this practice. The applicability of the absorbent polymer, adsorbent or combination to extract the compound(s) of interest must be demonstrated before use.

1.5 Where used it is the responsibility of the user to validate the application of SPME to the analytes of interest.

1.6 The values stated in SI units are to be regarded as the standard.

1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazard statements, see Section 9.

 
2. Referenced Documents

D4210-89(1996)e1

Standard Practice for Intralaboratory Quality Control Procedures and a Discussion on Reporting Low-Level Data (Withdrawn 2002)

D1129-24

Standard Terminology Relating to Water

D1193-24

Standard Specification for Reagent Water

D3856-11

Standard Guide for Management Systems in Laboratories Engaged in Analysis of Water

D3694-96(2024)

Standard Practices for Preparation of Sample Containers and for Preservation of Organic Constituents

D6520-18

Standard Practice for the Solid Phase Micro Extraction (SPME) of Water and its Headspace for the Analysis of Volatile and Semi-Volatile Organic Compounds