ASTM E1944-98(2021) - 1.4.2021
 
Significance and Use

4.1?This practice permits an analyst to compare the general performance of a laboratory instrument on any given day with the prior performance of that instrument. This practice is not intended for comparison of different instruments with each other, nor is it directly applicable to dedicated process FT-NIR analyzers. This practice requires the use of a check sample compatible with the instrument under test as described in 5.3.

 
1. Scope

1.1?This practice covers two levels of tests to measure the performance of laboratory Fourier transform near infrared (FT-NIR) spectrometers. This practice applies to the short-wave near infrared region, approximately 800 nm (12 500 cm1) to 1100 nm (9090.91 cm1); and the long-wavelength near infrared region, approximately 1100 nm (9090.91 cm1) to 2500 nm (4000 cm1). This practice is intended mainly for transmittance measurements of gases and liquids, although it is broadly applicable for reflectance measurements.

1.2?The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.3?This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.4?This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

 
2. Referenced Documents

E1421-99(2021)

Standard Practice for Describing and Measuring Performance of Fourier Transform Mid-Infrared (FT-MIR) Spectrometers: Level Zero and Level One Tests

E932-89(2021)

Standard Practice for Describing and Measuring Performance of Dispersive Infrared Spectrometers

E168-16(2023)

Standard Practices for General Techniques of Infrared Quantitative Analysis

E1252-98(2021)

Standard Practice for General Techniques for Obtaining Infrared Spectra for Qualitative Analysis

E131-10(2023)

Standard Terminology Relating to Molecular Spectroscopy