ASTM E827-07 - 1.10.2007
 
1. Scope

1.1 This practice outlines the necessary steps for the identification of elements in a given Auger spectrum obtained using conventional electron spectrometers. Spectra displayed as either the electron energy distribution (direct spectrum) or the first derivative of the electron energy distribution are considered.

1.2 This practice applies to Auger spectra generated by electron or X-ray bombardment of the specimen surface and can be extended to spectra generated by other methods such as ion bombardment.

1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

E673-03

Standard Terminology Relating to Surface Analysis (Withdrawn 2012)

E983-19

Standard Guide for Minimizing Unwanted Electron Beam Effects in Auger Electron Spectroscopy

E984-12(2020)

Standard Guide for Identifying Chemical Effects and Matrix Effects in Auger Electron Spectroscopy

E1523-15

Standard Guide to Charge Control and Charge Referencing Techniques in X-Ray Photoelectron Spectroscopy

ISO17974:2002

Surface Chemical AnalysisHigh-Resolution Auger Electron SpectrometersCalibration of Energy Scales for Elemental and Chemical-State Analysis