ASTM E925-83(1994)e1 - 10.3.2002
 
Significance and Use

This practice permits an analyst to compare the performance of an instrument to the manufacturer’supplied performance specifications and to verify its suitability for continued routine use. It also provides generation of calibration monitoring data on a periodic basis, forming a base from which any changes in the performance of the instrument will be evident.

 
1. Scope

1.1 This practice covers the parameters of spectrophotometric performance that are critical for testing the adequacy of instrumentation for most routine tests and methods within the wavelength range of 200 to 700 nm and the absorbance range of 0 to 2. The recommended tests provide a measurement of the important parameters controlling results in spectrophotometric methods, but it is specifically not to be inferred that all factors in instrument performance are measured.

1.2 This practice may be used as a significant test of the performance of instruments for which the spectral slit width does not exceed 2 nm and for which the manufacturer's specifications for wavelength and absorbance accuracy do not exceed the performance tolerances employed here. This practice employs an illustrative tolerance of ± 1 % relative for the error of the absorbance scale over the range of 0.2 to 2.0, and of 1.0 nm for the error of the wavelength scale. A suggested maximum stray radiant power ratio of 4 x 10-4 yields <1 % absorbance bias at an absorbance of 2. These tolerances are chosen to be compatible with many chemical applications while comfortably exceeding the uncertainty of the certified values for the reference materials and typical manufacturer's specifications for error in the wavelength and absorbance scales of the instrument under test. The user is encouraged to develop and use tolerance values more appropriate to the requirements of the end use application. This procedure is designed to verify quantitative performance on an ongoing basis and to compare one instrument's performance with that of other similar units. Refer to Practice E 275 to extensively evaluate the performance of an instrument.

1.3 This practice should be performed on a periodic basis, the frequency of which depends on the physical environment within which the instrumentation is used. Thus, units handled roughly or used under adverse conditions (exposed to dust, chemical vapors, vibrations, or combinations thereof) should be tested more frequently than those not exposed to such conditions. This practice should also be performed after any significant repairs are made on a unit, such as those involving the optics, detector, or radiant energy source.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

E275-08(2022)

Standard Practice for Describing and Measuring Performance of Ultraviolet and Visible Spectrophotometers

E387-04(2022)

Standard Test Method for Estimating Stray Radiant Power Ratio of Dispersive Spectrophotometers by the Opaque Filter Method

E1866-97(2021)

Standard Guide for Establishing Spectrophotometer Performance Tests

NIST?Special?Publication?260-54

Certification and Use of Acidic Potassium Dichromate Solutions As An Ultraviolet Absorbance Standard

NIST?Special?Publication?260-102

Holmium Oxide Solution Wavelength Standard from 240 to 640 nm--SRM 2034

E131-10(2023)

Standard Terminology Relating to Molecular Spectroscopy

NIST?Special?Publication?260-140

Technical Specifications for Certification of Spectrophotometric NTRMs

ISO?Guide?34

General Requirements for the Competence of Reference Material Producers

ISO?17025

General Requirements for the Competence of Testing and Calibration Laboratories

E169-16(2022)

Standard Practices for General Techniques of Ultraviolet-Visible Quantitative Analysis

NIST?Special?Publication?260-116

Glass Filters as a Standard Reference Material for Spectrophotometry--Selection, Preparation, Certification, and Use of SRM 930 and SRM 1930