ASTM F1440-92(1997) - 10.7.1997
 
1. Scope

1.1 This practice describes a method for the fatigue testing of metallic stemmed femoral components used in hip arthroplasty. The described method is intended to be used to evaluate the comparison of various designs and materials used for stemmed femoral components used in the arthroplasty. This practice covers procedures for the performance of fatigue tests using (as a forcing function) a periodic constant amplitude force.

1.2 This practice applies primarily to one-piece prostheses and modular components, with head in place such that prostheses should not have an anterior/posterior bow, and should have a nearly straight section on the distal 50 mm of the stem. This practice may require modifications to accommodate other femoral stem designs.

1.3 For additional information see Refs. (1) through (5).

 
2. Referenced Documents

E466-21

Standard Practice for Conducting Force Controlled Constant Amplitude Axial Fatigue Tests of Metallic Materials

E4-21

Standard Practices for Force Calibration and Verification of Testing Machines

E467-21

Standard Practice for Verification of Constant Amplitude Dynamic Forces in an Axial Fatigue Testing System

E468-18

Standard Practice for Presentation of Constant Amplitude Fatigue Test Results for Metallic Materials