ASTM F2582-20 - 15.11.2020
 
Significance and Use

5.1?This test method should be used to evaluate and compare different femoral and acetabular prosthesis designs to assess the damage tolerance under controlled laboratory conditions.

5.2?Although the methodology described attempts to identify physiologically relevant motions and loading conditions, the interpretation of results is limited to an in-vitro comparison between different femoral and acetabular prosthesis designs regarding their ability to resist impingement damage modes (defined in 8.2) under the stated test conditions.

 
1. Scope

1.1?This test method covers a procedure to simulate dynamic impingement between femoral and acetabular components in a hip replacement; the subsequent qualitative assessment of damage modes (as outlined in 8.2); and, if necessary, quantitative assessment of changes in modular component attachment strength.

1.2?This test method can be used to evaluate impingement between femoral components and the following: single-piece, modular, semi-constrained, bipolar, constrained, or dual mobility acetabular components, manufactured from polymeric, metallic, or ceramic materials.

1.3?The values stated in SI units are regarded as the standard.

1.4?This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.5?This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

 
2. Referenced Documents

F2003-02(2022)

Standard Practice for Accelerated Aging of Ultra-High Molecular Weight Polyethylene After Gamma Irradiation in Air

E467-21

Standard Practice for Verification of Constant Amplitude Dynamic Forces in an Axial Fatigue Testing System

E4-24

Standard Practices for Force Calibration and Verification of Testing Machines

F2091-15

Standard Specification for Acetabular Prostheses (Withdrawn 2023)

F2033-25

Standard Specification for Hip Joint Replacement Bearing Surfaces