ASTM E1835-14 - 1.10.2014
 
Significance and Use

5.1 This test method is used for the analysis of nickel alloy samples by FAAS to check compliance with compositional specifications. It is assumed that all who use the procedure will be trained analysts capable of performing common laboratory procedures skillfully and safely. It is expected that the work will be performed in a properly equipped laboratory and that proper waste disposal procedures will be followed. Appropriate quality control practices must be followed such as those described in Guide E882.

5.2 Interlaboratory Studies (ILS)5, 6International interlaboratory studies were conducted by ISO/TC 155/SC4, Analysis of nickel alloys. Results were evaluated in accordance with ISO 5725:1986 and restated to conform to Practice E1601. The method was published as ISO 7530, Parts 1 through 9. The published ISO statistics are summarized separately for each analyte to correspond with Practice E1601.

5.3 In this test method, some matrix modifiers are specified. However, other additives have come into common use since the original publication of this test method. These may be equally or more effective but have not been tested. It is the responsibility of the user to validate the use of such additives or the use of different dilutions, or both.

 
1. Scope

1.1 This test method covers analysis of nickel alloys by flame atomic absorption spectrometry (FAAS) for the following elements:

Element

Compostiton Range, %

 

 

Aluminum

0.2  to 4.0

Chromium

0.01 to 4.0

Cobalt

0.01 to 4.0

Copper

0.01 to 4.0

Iron

0.1  to 4.0

Manganese

0.1  to 4.0

Silicon

0.2  to 1.0

Vanadium

0.05 to 1.0

1.2 The composition ranges of these elements can be expanded by the use of appropriate standards.

1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazards associated with the use of this test method, see Practices E50 and the warning statements included in this test method.

 
2. Referenced Documents

ISO?7530

Parts 1 through 9--Nickel Alloys--Flame Atomic Absorption Spectrometric Analysis

ISO?5725:1986

Precision of Test Methods--Determination of Repeatability and Reproducibility for a Standard Test Method by Inter-laboratory Tests

E1812-96

Standard Practice for Optimization of Flame Atomic Absorption Spectrometric Equipment

E1601-19

Standard Practice for Conducting an Interlaboratory Study to Evaluate the Performance of an Analytical Method

E882-10(2016)e1

Standard Guide for Accountability and Quality Control in the Chemical Analysis Laboratory (Includes all amendments and changes 10/27/2017).

E135-23a

Standard Terminology Relating to Analytical Chemistry for Metals, Ores, and Related Materials (Includes all amendments and changes 7/17/2023).

D1193-06(2018)

Standard Specification for Reagent Water

E29-22

Standard Practice for Using Significant Digits in Test Data to Determine Conformance with Specifications

E50-17

Standard Practices for Apparatus, Reagents, and Safety Considerations for Chemical Analysis of Metals, Ores, and Related Materials