ASTM C1030-03 - 10.7.2003
 
Significance and Use

The determination of isotopic composition by gamma-ray spectrometry is a nondestructive technique and when used with other nondestructive techniques, such as calorimetry (Test Method C 1458) or neutron counting (Test Methods C 1207C 1493C 1500), can provide a totally nondestructive plutonium assay necessary for material accountancy and safeguards needs.

Since gamma-ray spectrometry systems are typically automated, the routine use of the test method is fast, reliable, and is not labor intensive. Since the test method is nondestructive, requiring no sample preparation, it does not create waste disposal problems.

This test method assumes that the isotopic composition of plutonium in the sample being measured is homogeneous (see see 7.2.4 and (5)).

The 242Pu abundance is not measured by this test method and must be estimated from isotopic correlation techniques, stream averages, historical information, or other measurement techniques.

A daughter product of 241Pu is 241Am. The 241Am/239Pu atom ratio can also be determined by means of this test method (assuming a homogeneous isotopic distribution of plutonium and 241Am) and is necessary for the correct interpretation of a calorimetric heat measurement.

The isotopic composition of a given batch or sample of plutonium is an attribute of that sample and, once determined, can be used in subsequent inventory measurements to verify the identity of a sample within the measurement uncertainties.

The method can also measure the ratio of other gamma emiting isotopes to plutonium assuming they have the same spatial distribution as the plutonium in the sample. Some of these “other” gamma-emitting isotopes include isotopes of uranium, neptium, curium, cesium, and other fission products. (The same methods of this standard can be used to measure the isotopic composition of uranium in samples containing only uranium (4–6)).

 
1. Scope

1.1 This test method is applicable to the determination of isotopic abundances in isotopically homogeneous Pu-bearing materials. This test method may be applicable to other plutonium-bearing materials, some of which may require modifications to the described test method.

1.2 The procedure is applicable to sample sizes ranging from a few tenths of a gram up to the maximum sample weight allowed by criticality limits.

1.3 Because 242 Pu has no useful gamma-ray signature, its isotopic abundance is not determined. Isotopic correlation techniques may be used to estimate its relative abundance (Refs 1, 2).

1.4 This test method has been demonstrated in routine use for isotopic abundances ranging from 94 to 70% 239 Pu. This test method has also been employed for isotopic abundances outside this range.

1.5 The values stated in SI units are to be regarded as the standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

C1500-08

Standard Test Method for Nondestructive Assay of Plutonium by Passive Neutron Multiplicity Counting

C1207-10(2018)

Standard Test Method for Nondestructive Assay of Plutonium in Scrap and Waste by Passive Neutron Coincidence Counting

C1316-08

Standard Test Method for Nondestructive Assay of Nuclear Material in Scrap and Waste by Passive-Active Neutron Counting Using 252Cf Shuffler

E181-23

Standard Guide for Detector Calibration and Analysis of Radionuclides in Radiation Metrology for Reactor Dosimetry

C698-24

Standard Test Methods for Chemical, Mass Spectrometric, and Spectrochemical Analysis of Nuclear-Grade Mixed Oxides ((U, Pu)O2)

C1458-24

Standard Test Method for Nondestructive Assay of Plutonium, Tritium and 241Am by Calorimetric Assay

E267-90(2001)

Standard Test Method for Uranium and Plutonium Concentrations and Isotopic Abundances (Withdrawn 2006)

C697-24

Standard Test Methods for Chemical, Mass Spectrometric, and Spectrochemical Analysis of Nuclear-Grade Plutonium Dioxide Powders and Pellets

C982-03

Standard Guide for Selecting Components for Energy-Dispersive X-Ray Fluorescence (XRF) Systems (Withdrawn 2008)

C1493-19

Standard Test Method for Non-Destructive Assay of Nuclear Material in Waste by Passive and Active Neutron Counting Using a Differential Die-Away System

ANSI N15.36