ASTM C1343-96 - 10.7.1996
 
1. Scope

1.1 This test method covers the steps necessary for the preparation and analysis by X-ray fluorescence (XRF) of oils and organic solutions containing uranium.

1.2 The procedure is valid for those solutions containing 20 to 2000 μg uranium/mL as presented to the spectrometer.

1.3 This test method requires the use of an appropriate internal standard. Care must be taken to ascertain that samples analyzed by this test method do not contain the internal standard or that this contamination, whenever present, has been corrected for mathematically. Such corrections are not addressed in this procedure. Care must be taken that the internal standard and sample medium are compatible; that is, samples must be miscible with tri-n-butyl phosphate (TBP) and must not remove the internal standard from solution. Alternatively, a scatter line may be used as the internal standard.

1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 9 and Note 2.

 
2. Referenced Documents

C1110-03

Standard Practice for Sample Preparation for X-Ray Emission Spectrometric Analysis of Uranium in Ores Using the Glass Fusion or Pressed Powder Method

C1254-26

Standard Test Method for Determination of Uranium in Mineral Acids by X-Ray Fluorescence

D1193-24

Standard Specification for Reagent Water

E135-26a

Standard Terminology Relating to Analytical Chemistry for Metals, Ores, and Related Materials (Includes all amendments and changes 8/6/2026).

ANSI/HPS N43.2–2001