ASTM C1832-23 - 1.1.2023
 
Significance and Use

5.1Uranium material is used as a fuel in certain types of nuclear reactors. To be suitable for use as nuclear fuel, the starting material shall meet certain specifications such as those described in Specifications C753, C776, C787, C833, C967, C996, and C1008, or as specified by the purchaser. The isotope amount ratios of uranium material can be measured by mass spectrometry following this test method to ensure that they meet the specification.

5.2The MTE method can be used for a wide range of sample sizes even in samples containing as low as 20 g of uranium. If the uranium sample is in the form of uranium hexafluoride, it has to be converted into a uranium nitrate solution for measurement by the MTE method. The concentration of the loading solution for MTE has to be in the range of 1 mg/g to 6 mg/g to allow a sample loading of 2 g to 6 g of uranium. A minimum loading of 3 g uranium per filament is strongly recommended. This is needed to have a sufficient and stable ion signal especially for the two minor isotopes (234U and 236U) thus enabling the internal calibration of SEM versus the Faraday cups using the 234U ion beam signal during the measurement.

5.3Until now, the instrument capabilities for the MTE method have only been implemented on the TRITON TIMS instrument.5 Therefore, all recommendations for measurement parameters in this test method are specified for the TRITON instrument. The manufacturers of other TIMS instruments (for example, IsotopX and Nu Instruments) have indicated plans to implement the modifications needed in their instruments to use the MTE method.

5.4The MTE method described here can also be extended to measurement of elements other than uranium. Note that the MTE method has already been implemented for plutonium and calcium.

 
1. Scope

1.1This test method describes the determination of the isotope amount ratios of uranium material as nitrate solutions by the modified total evaporation (MTE) method using a thermal ionization mass spectrometer (TIMS) instrument.

1.2The analytical performance in the determination of the 235U/238U major isotope amount ratio by MTE is similar to the (classical) total evaporation (TE) method as described in C1672. However, in the MTE method, the evaporation process is interrupted on a regular basis to allow measurements and subsequent corrections for background from peak tailing, perform internal calibration of a secondary electron multiplier (SEM) detector versus the Faraday cups, peak centering, and ion source refocusing. Performing these calibrations and corrections on a regular basis during the measurement, improves precision, and significantly reduces uncertainties for the minor isotope amount ratios 234U/238U and 236U/238U as compared to the TE method.

1.3In principle, the MTE method may yield major isotope amount ratios without the need for mass fractionation correction. However, depending on the measurement conditions, small variations are observed between sample turrets. Therefore, a small correction based on measurements of a certified reference material is recommended to improve consistency. The uncertainty around the mass fractionation correction factor usually includes unity.

1.4UnitsThe values stated in SI units are to be regarded as standard. The values given in parentheses after SI units are provided for information only and are not considered standard.

1.5This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.6This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

 
2. Referenced Documents

D1193-24

Standard Specification for Reagent Water

E2655-14(2020)

Standard Guide for Reporting Uncertainty of Test Results and Use of the Term Measurement Uncertainty in ASTM Test Methods

E2586-19(2024)

Standard Practice for Calculating and Using Basic Statistics

C833-23

Standard Specification for Sintered (Uranium-Plutonium) Dioxide Pellets for Light Water Reactors

C1672-23

Standard Test Method for Determination of the Uranium, Plutonium or Americium Isotopic Composition or Concentration by the Total Evaporation Method Using a Thermal Ionization Mass Spectrometer

C859-26

Standard Terminology Relating to Nuclear Materials

C996-20(2025)

Standard Specification for Uranium Hexafluoride Enriched to Less Than 5 % 235U

C1008-99

Standard Specification for Sintered (Uranium-Plutonium) Dioxide Pellets-Fast Reactor Fuel

C1068-21

Standard Guide for Qualification of Measurement Methods by a Laboratory Within the Nuclear Industry

C1128-23

Standard Guide for Preparation of Working Reference Materials for Use in Analysis of Nuclear Fuel Cycle Materials

C1156-26

Standard Guide for Establishing Calibration for a Measurement Method Used to Analyze Nuclear Fuel Cycle Materials

C1347-08(2023)

Standard Practice for Preparation and Dissolution of Uranium Materials for Analysis

C1411-25

Standard Practice for The Ion Exchange Separation of Uranium and Plutonium Prior to Isotopic Analysis

C967-25

Standard Specification for Uranium Ore Concentrate

C1625-24

Standard Test Method for Uranium and Plutonium Concentrations and Isotopic Abundances by Thermal Ionization Mass Spectrometry

C753-16a(2021)

Standard Specification for Nuclear-Grade, Sinterable Uranium Dioxide Powder (Includes all amendments and changes 10/22/2021).

C776-17(2022)

Standard Specification for Sintered Uranium Dioxide Pellets for Light Water Reactors

C787-20(2025)

Standard Specification for Uranium Hexafluoride for Enrichment