ASTM C696-11 - 1.9.2011
 
Significance and Use

Uranium dioxide is used as a nuclear-reactor fuel. In order to be suitable for this purpose, the material must meet certain criteria for uranium content, stoichiometry, isotopic composition, and impurity content. These test methods are designed to show whether or not a given material meets the specifications for these items as described in Specifications C753 and C776.

An assay is performed to determine whether the material has the minimum uranium content specified on a dry weight basis.

The stoichiometry of the oxide is useful for predicting its sintering behavior in the pellet production process.

Determination of the isotopic content of the uranium in the uranium dioxide powder is made to establish whether the effective fissile content is in compliance with the purchaser's specifications.

Impurity content is determined to ensure that the maximum concentration limit of certain impurity elements is not exceeded. Determination of impurities is also required for calculation of the equivalent boron content (EBC).

 
1. Scope

1.1 These test methods cover procedures for the chemical, mass spectrometric, and spectrochemical analysis of nuclear-grade uranium dioxide powders and pellets to determine compliance with specifications.

1.2 The analytical procedures appear in the following order:

Sections

Uranium by Ferrous Sulfate Reduction in Phosphoric Acid and Dichromate Titration Method

C1267 Test Method for Uranium By Iron (II) Reduction In Phosphoric Acid Followed By Chromium (VI) Titration In The Presence of Vanadium

Uranium and Oxygen Uranium Atomic Ratio by the Ignition (Gravimetric) Impurity Correction Method

C1453 Standard Test Method for the Determination of Uranium by Ignition and Oxygen to Uranium Ratio (O/U) Atomic Ratio of Nuclear Grade Uranium Dioxide Powders and Pellets

Carbon (Total) by Direct Combustion-Thermal Conductivity Method

C1408 Test Method for Carbon (Total) in Uranium Oxide Powders and Pellets By Direct Combustion-Infrared Detection Method

Total Chlorine and Fluorine by Pyrohydrolysis Ion-Selective Electrode Method

C1502 Standard Test Method for the Determination of Total Chlorine and Fluorine in Uranium Dioxide and Gadolinium Oxide    

Moisture by the Coulometric, Electrolytic Moisture Analyzer Method

7-14

Nitrogen by the Kjeldahl Method

15-22

Isotopic Uranium Composition by Multiple-Filament Surface Ionization Mass Spectrometric Method

Spectrochemical Determination of Trace Elements in High-Purity Uranium Dioxide

23-30

Silver, Spectrochemical Determination of, by Gallium OxideCarrier D-C Arc Technique

31 and 32

Rare Earths by Copper Spark-Spectrochemical Method

Impurity Elements by a Spark-Source Mass Spectrographic Method

C761 Test Method for Chemical, Mass Spectrometric, Spectrochemical, Nuclear, and Radiochemical Analysis of Uranium Hexafluoride

C1287 Test Method for Determination of Impurities In Uranium Dioxide By Inductively Coupled Plasma Mass Spectrometry

Surface Area by Nitrogen Absorption Method

33-39

Total Gas in Reactor-Grade Uranium Dioxide Pellets

Thorium and Rare Earth Elements by Spectroscopy

Hydrogen by Inert Gas Fusion

C1457 Standard Test Method for Determination of Total Hydrogen Content of Uranium Oxide Powders and Pellets by     Carrier Gas Extraction    

Uranium Isotopic Analysis by Mass Spectrometry

C1413 Test Method for Isotopic Analysis of Hydrolysed Uranium Hexafluoride and Uranyl Nitrate Solutions By Thermal Ionization Mass Spectrometry

7.1 This test method covers the determination of moisture in uranium dioxide samples. Detection limits are as low as 10 μg.

15.1 This test method covers the determination of nitride nitrogen in uranium dioxide in the range from 10 to 250 μg.

23.1 This test method covers the spectrographic analysis of nuclear-grade UO2 for the 26 elements in the ranges indicated in Table 1.

23.2 For simultaneous determination of trace elements by plasma emission spectroscopy refer to Test Method C761.

TABLE 1 Recommended Analytical Spectral Lines and Concentration Range of Trace Elements

ElementAnalytical Line,
°AA
Concentration range, μg/g of U

AgB

3280.68

0.1 to 50

Al

2367.06

10 to 200

As

2349.84

5 to 50

B

2497.73

0.10 to 5

Ba

4554.04

10 to 300

Be

2348.61

0.1 to 3

Bi

3067.72

5 to 50

Ca

4226.73

1 to 50

Cd

2288.02

0.15 to 5

Co

3453.51

5 to 50

Cr

2843.25

10 to 100

Cu

3247.54

1 to 10

Fe

2462.64

10 to 300

In

3256.09

5 to 50

Mg

2779.83

10 to 100

Mn

2605.69

5 to 50

Mo

3132.59

0.5 to 10

Na

3302.32

80 to 400

Ni

3050.82

5 to 100

P

2553.28

50 to 500

Pb

2833.07

5 to 50

Sb

2598.05

1 to 50

Si

2435.16

10 to 200

Sn

3175.02

5 to 50

Ti

3361.26

1 to 100

V

3183.41

1 to 100

Zn

3345.02

20 to 300

Zr

3438.23

25 to 300

A All of the above lines are photographed in the second order, except barium and calcium which are first order lines.

B A gallium oxide carrier must be used for silver. See Test Method E402.

31.1 This test method covers the spectrochemical determination of silver in nuclear-grade uranium dioxide. The relative standard deviation is 15 % for the concentration range of 0.1 to 50 μg/g.

33.1 This procedure is designed for a rapid determination of surface area of nuclear-grade uranium dioxide (UO2) powders. The determination of surface area by this procedure is automatic, simple, and fast enough to be used in quality control work. The range of analysis is from 1 to 1500 m2/g.

 
2. Referenced Documents

C1408-24

Standard Test Method for Carbon (Total) in Uranium Oxide Powders and Pellets By Direct Combustion-Infrared Detection Method

C1413-26

Standard Test Method for Isotopic Analysis of Hydrolyzed Uranium Hexafluoride and Uranyl Nitrate Solutions by Thermal Ionization Mass Spectrometry

E402-02

Standard Test Method for Spectrographic Analysis of Uranium Oxide (U3O8)by Gallium Oxide-Carrier Technique (Withdrawn 2007)

E130-08

Standard Practice for Designation of Shapes and Sizes of Graphite Electrodes (Withdrawn 2013)

E115-97

Standard Practice for Photographic Processing in Optical Emission Spectrographic Analysis (Withdrawn 2002)

D1193-24

Standard Specification for Reagent Water

C1287-18

Standard Test Method for Determination of Impurities in Nuclear Grade Uranium Compounds by Inductively Coupled Plasma Mass Spectrometry

C753-16a(2021)

Standard Specification for Nuclear-Grade, Sinterable Uranium Dioxide Powder (Includes all amendments and changes 10/22/2021).

C1457-26

Standard Test Method for Determination of Total Hydrogen Content of Uranium Oxide Powders and Pellets by Carrier Gas Extraction

C1502-24

Standard Test Method for Determination of Total Chlorine and Fluorine in Uranium Dioxide and Gadolinium Oxide

C1347-08(2023)

Standard Practice for Preparation and Dissolution of Uranium Materials for Analysis

C776-17(2022)

Standard Specification for Sintered Uranium Dioxide Pellets for Light Water Reactors

C1453-19

Standard Test Method for the Determination of Uranium by Ignition and the Oxygen to Uranium (O/U) Atomic Ratio of Nuclear Grade Uranium Dioxide Powders and Pellets

C1267-17(2022)

Standard Test Method for Uranium by Iron (II) Reduction in Phosphoric Acid Followed by Chromium (VI) Titration in the Presence of Vanadium

C761-24

Standard Test Methods for Chemical, Mass Spectrometric, Spectrochemical, Nuclear, and Radiochemical Analysis of Uranium Hexafluoride