ASTM C697-10 - 1.6.2010
 
Significance and Use

Plutonium dioxide is used in mixtures with uranium dioxide as a nuclear-reactor fuel. In order to be suitable for this purpose, the material must meet certain criteria for plutonium content, isotopic composition, and impurity content. These test methods are designed to show whether or not a given material meets the specifications for these items as described in Specification C757.

An assay is performed to determine whether the material has the minimum plutonium content specified on a dry weight basis.

Determination of the isotopic content of the plutonium in the plutonium dioxide powder is made to establish whether the effective fissile content is in compliance with the purchaser's specifications.

Impurity content is determined to ensure that the maximum concentration limit of certain impurity elements is not exceeded. Determination of impurities is also required for calculation of the equivalent boron content (EBC).

 
1. Scope

1.1 These test methods cover procedures for the chemical, mass spectrometric, and spectrochemical analysis of nuclear-grade plutonium dioxide powders and pellets to determine compliance with specifications.

1.2 The analytical procedures appear in the following order:

Sections

Plutonium Sample Handling

8 to 10

Plutonium by Controlled-Potential Coulometry

Plutonium by Ceric Sulfate Titration

Plutonium by Amperometric Titration with Iron(II)

Plutonium by Diode Array   Spectrophotometry

Nitrogen by Distillation Spectrophotometry Using Nessler
Reagent

11 to 18

Carbon (Total) by Direct Combustion–Thermal Conductivity

19 to 30

Total Chlorine and Fluorine by Pyrohydrolysis

31 to 38

Sulfur by Distillation Spectrophotometry

39 to 47

Plutonium Isotopic Analysis by Mass Spectrometry

Rare Earth Elements by Spectroscopy

48 to 55

Trace Elements by Carrier–Distillation Spectroscopy

56 to 63

Impurities by ICP-AES

Impurity Elements by Spark-Source Mass Spectrography

64 to 70

Moisture by the Coulometric Electrolytic Moisture Analyzer

71 to 78

Total Gas in Reactor-Grade Plutonium Dioxide Pellets

79 to 86

Plutonium-238 Isotopic Abundance by Alpha Spectrometry

Americium-241 in Plutonium by Gamma-Ray Spectrometry

Rare Earths By Copper Spark-Spectroscopy

87 to 96

Plutonium Isotopic Analysis by Mass Spectrometry

97 to 105

Oxygen-To-Metal Atom Ratio by Gravimetry

106 to 114

1.3 The values stated in SI units are to be regarded as standard. The values given in parentheses are for information only.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific precautionary statements, see Sections 6, 15, 24, 111, and 52.9 and 101.5.1 .

8.1 This test method covers the conditions necessary to preserve the integrity of plutonium dioxide samples. Conditions listed here are directed toward the analytical chemist. However, they are just as applicable to any group handling the material.

11.1 This test method covers the determination of 5 to 100 μg/g of nitride nitrogen in 1-g samples of nuclear-grade plutonium dioxide.

19.1 This test method covers the determination of 10 to 200 μg of residual carbon in nuclear-grade plutonium dioxide.

31.1 This test method covers the determination of 5 to 100 μg/g of chlorine and 1 to 100 μg/g of fluorine in 1-g samples of nuclear-grade plutonium dioxide.

39.1 This test method coves the determination of sulfur in the concentration range from 10 to 600 μg/g for samples of nuclear-grade plutonium dioxide powders or pellets.

48.1 This test method covers the determination of dysprosium, europium, gadolinium, and samarium in plutonium dioxide (PuO2) in concentrations of 0.1 to 10 μg/g of PuO2.


FIG. 5 Schematic Diagram of Atmosphere Chamber

56.1 This test method covers the determination of 36 impurity elements in nuclear-grade plutonium dioxide (PuO2). The concentration ranges covered by the method and the analytical lines used for each element are listed in Table 1.

TABLE 1 Impurity Elements, Lines, and Concentration Limits

ElementWavelength, ÅConcentration
Range, ppm

Al

2567.99A

10–500

Al

3082.68A

Ag

3280.68

1–250

As

2349.84

10–250

Au

2427.95

2.5–50

B

2497.73

1–50

Ba

4554.03

2.5–100

(first order)

Be

2348.61

0.5–25

Bi

3067.72

2.5–100

Ca

3933.67

10–1000

Cd

2288.02

1–100

Co

2407.25

10–1000

Co

3044.00

Cu

3247.54

1–250

Cr

2835.63A

5–500

Fe

2983.57A

10–1000

Ga

2450.07A

1–500

Ga

2943.64

In

3039.36

2.5–100

K

7664.91

3–1000

(first order)

Mg

2802.70

1–1000

Mn

2798.27

5–250

MoB

3132.59

5–250

Ni

3002.49A

10–1000

P

2553.28

100–1000

Pb

2614.18

5–100

Pd

3242.70

2.5–100

Sb

2598.05

2.5–100

Si

2514.32A

10–300

Sn

2863.33

2.5–100

Sr

4607.33

5–100

(first order)

Ti

3223.52

5–100

VB

3183.98A

5–100

WB

2724.35A

20–200

Zn

3345.02

10–250

Zr

3438.23

5–100

A These elements are determined with the aid of a microphotometer.

B These elements are analyzed by using a AgCl carrier and PdCl2 as an internal standard. Mix 20 mg of PdCl2 with 10.0 g of AgCl. Add 80 mg of this mixture to 120 mg of PuO2 sample-blend before loading 50-mg charges in a graphite electrode ASTM-Type-S-4 (follow 96.2.2).

71.1 This test method covers the determination of moisture in plutonium dioxide samples. Detection limits are as low as 10 μg.

79.1 This test method covers the measurement of volatiles other than water in reactor-grade plutonium dioxide pellets (16). The lower limit of the test method using the described equipment and a sample of approximately 1 g is 0.01 cm3 per gram of sample at STP conditions.

87.1 This test method covers the determination of rare earths in plutonium dioxide over the range from 10 to 200 μg/g.

97.1 This test method covers the determination of the isotopic composition of plutonium in nuclear-grade plutonium dioxide powders and pellets.

106.1 This test method is applicable to the determination of the oxygen-to-metal atom ratio (O/M) in plutonium dioxide pellets and powders.

 
2. Referenced Documents

E115-97

Standard Practice for Photographic Processing in Optical Emission Spectrographic Analysis (Withdrawn 2002)

E60-11(2022)e1

Standard Practice for Analysis of Metals, Ores, and Related Materials by Spectrophotometry (Includes all amendments and changes 6/29/2022).

D4327-17(2025)

Standard Test Method for Anions in Water by Suppressed Ion Chromatography

E130-08

Standard Practice for Designation of Shapes and Sizes of Graphite Electrodes (Withdrawn 2013)

C1415-24

Standard Test Method for 238Pu Isotopic Abundance by Alpha Spectrometry

D1193-24

Standard Specification for Reagent Water

C1672-23

Standard Test Method for Determination of the Uranium, Plutonium or Americium Isotopic Composition or Concentration by the Total Evaporation Method Using a Thermal Ionization Mass Spectrometer

C1637-21

Standard Test Method for Determination of Impurities in Plutonium Materials—Acid Digestion and Inductively Coupled Plasma-Mass Spectroscopy (ICP-MS) Analysis

C1625-24

Standard Test Method for Uranium and Plutonium Concentrations and Isotopic Abundances by Thermal Ionization Mass Spectrometry

C1432-23

Standard Test Method for Determination of Impurities in Plutonium: Acid Dissolution, Ion Exchange Matrix Separation, and Inductively Coupled Plasma-Atomic Emission Spectroscopic (ICP/AES) Analysis

C1068-21

Standard Guide for Qualification of Measurement Methods by a Laboratory Within the Nuclear Industry

C859-26

Standard Terminology Relating to Nuclear Materials

C852-09

Standard Guide for Design Criteria for Plutonium Gloveboxes

C757-16(2021)

Standard Specification for Nuclear-Grade Plutonium Dioxide Powder for Light Water Reactors

C1307-21

Standard Test Method for Plutonium Assay by Plutonium (III) Diode Array Spectrophotometry

E116-97

Standard Practice for Photographic Photometry in Spectrochemical Analysis (Withdrawn 2002)

C1235-99

Standard Test Method for Plutonium by Titanium(III)/Cerium(IV) Titration (Withdrawn 2005)

C1233-15(2021)

Standard Practice for Determining Equivalent Boron Contents of Nuclear Materials

C1206-02

Standard Test Method for Plutonium by Iron (II)/Chromium (VI) Amperometric Titration

C1168-23

Standard Practice for Preparation and Dissolution of Plutonium Materials for Analysis

C1165-23

Standard Test Method for Determining Plutonium by Controlled-Potential Coulometry in H2SO4 at a Platinum Working Electrode

C1108-23

Standard Test Method for Plutonium by Controlled-Potential Coulometry

C1268-23

Standard Test Method for Quantitative Determination of 241Am in Plutonium by Gamma-Ray Spectrometry