ASTM E2510-07 - 1.2.2007
 
Significance and Use

The test method calibrates or demonstrates conformity of the torque signal of a rheometer at ambient temperature.

A calibration factor thus determined may be used to obtain correct torque values.

This test method may be used in research, development, specification acceptance, and quality control or assurance.

 
1. Scope

1.1 This test method describes the calibration or performance conformance for the torque signal generated by commercial or custom-built rheometers. The specific range of the test depends upon the torque range of the rheometer.

1.2 SI units are the standard.

1.3 There is no ISO standard equivalent to this test method.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

E4-21

Standard Practices for Force Calibration and Verification of Testing Machines

E473-23b

Standard Terminology Relating to Thermal Analysis and Rheology (Includes all amendments and changes 10/5/2023).

E1142-23b

Standard Terminology Relating to Thermophysical Properties (Includes all amendments and changes 10/5/2023).

E617-23

Standard Specification for Laboratory Weights and Precision Mass Standards