ASTM D4222-98 - 10.3.1998
 
1. Scope

1.1 This test method covers the determination of nitrogen adsorption and desorption isotherms of catalysts and catalyst carriers at the boiling point of liquid nitrogen. A static volumetric measuring system is used to obtain about thirty equilibrium adsorption points on each branch of the isotherm, these points being evenly spread over the isotherm.

1.2 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

E177-26a

Standard Practice for Use of the Terms Precision and Bias in ASTM Test Methods (Includes all amendments and changes 5/14/2026).

E691-23

Standard Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method

D3766-24a

Standard Terminology Relating to Catalysts and Catalysis (Includes all amendments and changes 3/6/2024).

D3663-25

Standard Test Method for Surface Area of Catalysts and Catalyst Carriers

E456-13a(2022)e1

Standard Terminology Relating to Quality and Statistics (Includes all amendments and changes 4/16/2024).