ASTM E1184-98 - 10.5.1998
 
1. Scope

1.1 This practice covers a procedure for the determination of microgram per millilitre ([mu]g/mL) or lower concentrations of elements in solution using an electrothermal atomization device attached to an atomic absorption spectrophotometer. A general description of the equipment is provided. Recommendations are made for preparing the instrument for measurements, establishing optimum temperature conditions and other criteria which should result in determining a useful calibration concentration range, and measuring and calculating the test solution analyte concentration.

1.2 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific safety hazard statements are given in Section 9.

Note 1-This practice is a companion to Practice E663.
 
2. Referenced Documents

E50-17

Standard Practices for Apparatus, Reagents, and Safety Considerations for Chemical Analysis of Metals, Ores, and Related Materials

E131-10(2023)

Standard Terminology Relating to Molecular Spectroscopy

D1193-06(2018)

Standard Specification for Reagent Water

E135-23a

Standard Terminology Relating to Analytical Chemistry for Metals, Ores, and Related Materials (Includes all amendments and changes 7/17/2023).

E406-19

Standard Practice for Using Controlled Atmospheres in Atomic Emission Spectrometry