ASTM E1275-98e1 - 10.6.1999
 
1. Scope

1.1 This practice covers the handling, testing, and procedure for using a radiochromic film dosimetry system to measure absorbed dose in materials irradiated by photons or electrons in terms of absorbed dose in water.

1.2 This practice applies to radiochromic film dosimeters that can be used within part or all of the specified ranges as follows:

1.2.1 The absorbed dose range is 1 X 10-2 to 1 X 102 kGy.

1.2.2 The absorbed dose rate is 1 X 10-2 to 1 X 1013 Gy/s (1-4).

1.2.3 The radiation energy range for both photons and electrons is 0.1 to 50 MeV.

1.2.4 The irradiation temperature range is -78 to +60°C.

1.3 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

51707-15

Standard Guide for Estimation of Measurement Uncertainty in Dosimetry for Radiation Processing

51261-13(2020)

Standard Practice for Calibration of Routine Dosimetry Systems for Radiation Processing

E275-08(2022)

Standard Practice for Describing and Measuring Performance of Ultraviolet and Visible Spectrophotometers

E170-23

Standard Terminology Relating to Radiation Measurements and Dosimetry

E2628-09

Standard Practice for Dosimetry for Radiation Processing

ICRU Report 80

Dosimetry Systems for Use in Radiation Processing

ICRU Report 85a

Fundamental Quantities and Units for Ionizing Radiation

E2701-09

Standard Guide for Performance Characterization of Dosimeters and Dosimetry Systems for Use in Radiation Processing (Withdrawn 2014)