ASTM E1276-96e1 - 10.6.1999
 
1. Scope

1.1 This practice covers procedures for using hermetically-sealed polymethylmethacrylate (PMMA) dosimeters for measuring absorbed dose in materials irradiated by photons or electrons in terms of absorbed dose in water.

1.2 This practice covers systems that permit absorbed dose measurements under the following conditions:

1.2.1 The absorbed dose range is 0.1 to 50 kGy.

1.2.2 The absorbed dose rate is 1 X 10-2 to 1 X 107 Gy[dot]s-1.

1.2.3 The radiation energy range for photons is 0.1 to 50 MeV, and for electrons 3 to 50 MeV.

1.2.4 The irradiation temperature is -78 to +50°C.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

51707-15

Standard Guide for Estimation of Measurement Uncertainty in Dosimetry for Radiation Processing

51261-13(2020)

Standard Practice for Calibration of Routine Dosimetry Systems for Radiation Processing

E2701-09

Standard Guide for Performance Characterization of Dosimeters and Dosimetry Systems for Use in Radiation Processing (Withdrawn 2014)

E2628-09

Standard Practice for Dosimetry for Radiation Processing

E275-08(2022)

Standard Practice for Describing and Measuring Performance of Ultraviolet and Visible Spectrophotometers

E170-23

Standard Terminology Relating to Radiation Measurements and Dosimetry

VIM

International Vocabulary of Basic and General Terms in Metrology, 2008

GUM

Guide to the Expression of Uncertainty in Measurement, 1995.

ICRU Report 80

Dosimetry Systems for Use in Radiation Processing

ICRU Report 60

Fundamental Quantities and Units for Ionizing Radiation