ASTM E1401-96e1 - 10.6.1999
 
1. Scope

1.1 This practice covers the preparation, testing, and procedure for using the acidic aqueous silver dichromate dosimetry system to measure absorbed dose in water when exposed to ionizing radiation. The system consists of a dosimeter and appropriate analytical instrumentation. For simplicity, the system will be referred to as the dichromate system. It is classified as a reference standard dosimetry system (see Guide E 1261).

1.2 The practice describes the spectrophotometric analysis procedures for the dichromate system.

1.3 This practice applies only to [gamma] rays, x rays, and high-energy electrons.

1.4 This practice applies provided the following conditions are satisfied:

1.4.1 The absorbed dose range is from 2 X 10 to 5 X 10 Gy.

1.4.2 The absorbed dose rate does not exceed 600 Gy/ pulse with a pulse repetition rate not to exceed 12.5 Hz, or does not exceed an equivalent dose rate of 7.5 kGy/s from continuous sources (1).

1.4.3 For radionuclide gamma-ray sources, the initial photon energy shall be greater than 0.6 MeV. For bremsstrahlung photons, the initial energy of the electrons used to produce the bremsstrahlung photons shall be equal to or greater than 2 MeV. For electron beams, the initial electron energy shall be greater than 8 MeV.

Note 1-The lower energy limits given are appropriate for a cylindrical dosimeter ampoule of 12 mm diameter. Corrections for dose gradients across an ampoule of that diameter or less are not required. The dichromate system may be used at lower energies by employing thinner (in the beam direction) dosimeter containers (see ICRU Report 35).

1.4.4 The irradiation temperature of the dosimeter shall be above 0°C and should be below 80°C.

Note 2-The temperature coefficient of dosimeter response is known only in the range of 20° to 50°C (see 10.1.8). Use outside this range is not recommended.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific hazard statements are given in Note 6.

 
2. Referenced Documents

E668-20

Standard Practice for Application of Thermoluminescence-Dosimetry (TLD) Systems for Determining Absorbed Dose in Radiation-Hardness Testing of Electronic Devices

E925-09(2022)

Standard Practice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidth does not Exceed 2 nm

E958-13(2021)

Standard Practice for Estimation of the Spectral Bandwidth of Ultraviolet-Visible Spectrophotometers

E170-23

Standard Terminology Relating to Radiation Measurements and Dosimetry

E178-21

Standard Practice for Dealing With Outlying Observations

E275-08(2022)

Standard Practice for Describing and Measuring Performance of Ultraviolet and Visible Spectrophotometers

51261-13(2020)

Standard Practice for Calibration of Routine Dosimetry Systems for Radiation Processing

51707-15

Standard Guide for Estimation of Measurement Uncertainty in Dosimetry for Radiation Processing

52628-20

Standard Practice for Dosimetry in Radiation Processing

52701-13(2020)

Standard Guide for Performance Characterization of Dosimeters and Dosimetry Systems for Use in Radiation Processing

17025

General Requirements for the Competence of Testing and Calibration Laboratories

JCGM 100:2008, GUM

1995, with minor corrections, Evaluation of measurement data - Guide to the Expression of Uncertainty in Measurement

ICRU Report 35

Radiation Dosimetry: Electrons With Initial Energies Between 1 and 50 MeV

ICRU Report 80

Dosimetry Systems for Use in Radiation Processing

ICRU Report 85a

Fundamental Quantities and Units for Ionizing Radiation

E666-21

Standard Practice for Calculating Absorbed Dose From Gamma or X Radiation