ASTM E1856-13 - 15.11.2013
 
Significance and Use

5.1 This guide is recommended to be used by anyone acquiring data from a universal testing machine using a computerized data acquisition system.

 
1. Scope

1.1 This guide is intended to assist the user in the evaluation and documentation of computerized data acquisition systems used to acquire data from quasi-static tests, performed on universal testing machines. The report produced will aid in the correct use and calibration of the computerized universal testing machine.

1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

E8/E8M-24

Standard Test Methods for Tension Testing of Metallic Materials

E1012-19

Standard Practice for Verification of Testing Frame and Specimen Alignment Under Tensile and Compressive Axial Force Application

E691-23

Standard Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method

E6-23a

Standard Terminology Relating to Methods of Mechanical Testing (Includes all amendments and changes 3/29/2023).

E83-23

Standard Practice for Verification and Classification of Extensometer Systems

E74-18e1

Standard Practices for Calibration and Verification for Force-Measuring Instruments (Includes all amendments and changes 5/1/2019).

E4-21

Standard Practices for Force Calibration and Verification of Testing Machines