ASTM E1952-98 - 10.8.2001
 
1. Scope

1.1 This test method covers the determination of thermal conductivity of homogeneous, non-porous solid materials in the range of 0.10 to 1.0 W/(K*m) by modulated temperature differential scanning calorimeter. This range includes many polymeric, glass, and ceramic materials. Thermal diffusivity, which is elated to thermal conductivity through specific heat capacity and density, may also be derived. Thermal conductivity and diffusivity can be determined at one or more temperatures over the range of 0 to 90°C.

1.2 Electronic instrumentation or automated data analysis and reduction systems or treatments equivalent to this test method may be used.

1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are provided for information purposes only.

1.4 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

E968-02

Standard Practice for Heat Flow Calibration of Differential Scanning Calorimeters

E473-23b

Standard Terminology Relating to Thermal Analysis and Rheology (Includes all amendments and changes 10/5/2023).

E967-18

Standard Test Method for Temperature Calibration of Differential Scanning Calorimeters and Differential Thermal Analyzers

E1142-23b

Standard Terminology Relating to Thermophysical Properties (Includes all amendments and changes 10/5/2023).

E1231-19

Standard Practice for Calculation of Hazard Potential Figures of Merit for Thermally Unstable Materials

E2161-23c

Standard Terminology Relating to Performance Validation in Thermal Analysis and Rheology (Includes all amendments and changes 8/17/2023).