ASTM E2465-19 - 1.10.2019
 
Significance and Use

5.1 This procedure is suitable for manufacturing control and for verifying that the product meets specifications. It provides rapid, multi-element determinations with sufficient accuracy to assure product quality. The analytical performance data included may be used as a benchmark to determine if similar X-ray spectrometers provide equivalent precision and accuracy, or if the performance of a particular spectrometer has changed.

 
1. Scope

1.1 This test method covers the analysis of Ni-base alloys by wavelength dispersive X-ray Fluorescence Spectrometry for the determination of the following elements:

Element

Composition Range

Manganese

0.06 % to 1.6 %

Phosphorus

0.008 % to 0.015 %

Silicon

0.08 % to 0.6 %

Chromium

1.6 % to 22 %

Nickel

23 % to 77 %

Aluminum

0.20 % to 1.3 %

Molybdenum

0.03 % to 10 %

Copper

0.007 % to 2.5 %

Titanium

0.11 % to 3.0 %

Niobium

0.55 % to 5.3 %

Iron

0.17 % to 46 %

Tungsten

0.06 % to 0.50 %

Cobalt

0.04 % to 0.35 %

Note 1: Unless exceptions are noted, ranges can be extended by the use of suitable reference materials. Once these element ranges are extended they must be verified by some experimental means. This could include but not limited to Gage Repeatability and Reproducibility studies, Interlaboratory Round Robin studies, or both. Once these studies are completed, they will satisfy the ISO/IEC 17025 requirements for capability.

1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

 
2. Referenced Documents

E1601-19

Standard Practice for Conducting an Interlaboratory Study to Evaluate the Performance of an Analytical Method

E1622-94(1999)e1

Standard Practice for Correction of Spectral Line Overlap in Wavelength-Dispersive X-Ray Spectrometry (Withdrawn 2006)

E1361-02(2021)

Standard Guide for Correction of Interelement Effects in X-Ray Spectrometric Analysis

E305-21

Standard Practice for Establishing and Controlling Spark Atomic Emission Spectrochemical Analytical Curves

E135-23a

Standard Terminology Relating to Analytical Chemistry for Metals, Ores, and Related Materials (Includes all amendments and changes 7/17/2023).

ISO?17025