ASTM E2524-08 - 1.2.2008
 
Significance and Use
p>This test method is one of a series of tests listed in Practice F 748 and ISO 10993-4 to assess the biocompatibility of materials contacting blood in medical applications.

This test method is similar to Practice F 756 but modified to accommodate nanoparticulate materials.

 
1. Scope

1.1 This test method covers assessing the effect of nanoparticulate materials on the integrity of red blood cells.

1.2 This test method uses diluted whole blood incubated with nanoparticulate material and the hemoglobin released from damaged red blood cells is determined.

1.3 This test method is similar to Practice F 756 with the volumes reduced to accommodate nanoparticulate material.

1.4 This test method is part of the in-vitro preclinical characterization and is important for nanoparticulate material that will contact the blood in medical applications.

1.5 The values given in SI units are to be considered as the standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

F756-17

Standard Practice for Assessment of Hemolytic Properties of Materials

F1877-16

Standard Practice for Characterization of Particles

ISO 10993-4

Biological Evaluation of Medical Devices Part 4: Selection of Tests for Interactions with Blood

F1903-18

Standard Practice for Testing for Cellular Responses to Particles in vitro

F748-16

Standard Practice for Selecting Generic Biological Test Methods for Materials and Devices