1. Scope

1.1 This practice covers a procedure for checking some of the operating characteristics of an X-ray photoelectron spectrometer. Tests herein provide checks of the following instrument characteristics: X-ray photoelectron spectroscopy (XPS) signal intensity, background, energy resolution, short-term voltage stability, transmission, and energy scale linearity. It is meant for spectrometers with digital storage of counts in energy channels.

1.2 Limitations -This practice is meant to augment, and not to replace, the calibration procedures recommended by the manufacturer of the spectrometer. This practice is also not meant to be used as a means of comparison between X-ray photoelectron spectrometers, but only as a self-consistent check of the operating characteristics of an individual spectrometer.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

E673-03

Standard Terminology Relating to Surface Analysis (Withdrawn 2012)

ISO 24237

Surface Chemical Analysis-- X-ray Photoelectron Spectroscopy - Repeatability and Constancy of Intensity Scale

E2108-16

Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer

E1078-14(2020)

Standard Guide for Specimen Preparation and Mounting in Surface Analysis

E996-19

Standard Practice for Reporting Data in Auger Electron Spectroscopy and X-ray Photoelectron Spectroscopy