ASTM F1264-00 - 10.12.2001
 
1. Scope

1.1 This specification is intended to provide a characterization of the design and mechanical function of intramedullary fixation devices (IMFDs) specify labeling and material requirements, provide test methods for characterization of IMFD mechanical properties and identify needs for further development of test methods and performance criteria. The ultimate goal is to develop a standard which defines performance criteria and methods for measurement of performance-related mechanical characteristics of IMFDs and their fixation to bone. It is not the intention of this specification to define levels of performance or case-specific clinical performance of these devices, as insufficient knowledge is available to predict the consequences of the use of any of these devices in individual patients for specific activities of daily living. It is not the intention of this specification to describe or specify specific designs for IMFDs.

1.2 This specification describes IMFDs for surgical fixation of the skeletal system. It provides basic IFMD geometrical definitions, dimensions, classification, and terminology; labeling and material specifications; performance definitions; test methods and characteristics determined to be important to in-vivo performance of the device.

1.3 This specification includes four standard test methods;

1.3.1 Static Four-Point Bend Test Method—Annex A1 and

1.3.2 Static Torsion Test Method—Annex A2.

1.3.3 Bending Fatigue Test Method—Annex A3.

1.3.4 Test Method for Bending Fatigue of IMFD Locking Screws—Annex A4.

1.4 A rationale is given in Appendix X1.

1.5 The values stated in SI units are to be regarded as the standard.

 
2. Referenced Documents

F339-71(1993)

Specification for Cloverleaf Intramedullary Pins (Withdrawn 1998)

F543-23

Standard Specification and Test Methods for Metallic Medical Bone Screws

E1823-23

Standard Terminology Relating to Fatigue and Fracture Testing

E467-21

Standard Practice for Verification of Constant Amplitude Dynamic Forces in an Axial Fatigue Testing System

F2503-23e1

Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment (Includes all amendments and changes 10/24/2023).

F1611-20

Standard Specification for Intramedullary Reamers

F565-21

Standard Practice for Care and Handling of Orthopedic Implants and Instruments

F383

Practice for Static Bend and Torsion Testing of Intramedullary Rods

F2809-10

Standard Terminology Relating to Medical and Surgical Materials and Devices (Withdrawn 2019)

SAE J524

Seamless Low-Carbon Steel Tubing Annealed for Bending and Flaring

AMS 5050

E467-21

Standard Practice for Verification of Constant Amplitude Dynamic Forces in an Axial Fatigue Testing System

E1823-23

Standard Terminology Relating to Fatigue and Fracture Testing

F86-21

Standard Practice for Surface Preparation and Marking of Metallic Surgical Implants

F138-19

Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673)

A214/A214M-19

Standard Specification for Electric-Resistance-Welded Carbon Steel Heat-Exchanger and Condenser Tubes

A450/A450M-23

Standard Specification for General Requirements for Carbon and Low Alloy Steel Tubes

D790-17

Standard Test Methods for Flexural Properties of Unreinforced and Reinforced Plastics and Electrical Insulating Materials

E4-21

Standard Practices for Force Calibration and Verification of Testing Machines

E691-23

Standard Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method