ASTM F1611-00(2018) - 1.2.2018
 
1. Scope

1.1 This specification provides requirements for material, dimensions and tolerances, finish and marking, and care and handling for reamers intended to cut a cylindrical path along the medullary canal of diaphyseal bone.

1.2 Intramedullary reamers are commonly used to prepare the medullary canal for the insertion of intramedullary fixation devices (IMFD). As such, the relationship between the intramedullary reamer diameter and the IMFD's diameter are considered.

1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

 
2. Referenced Documents

A564/A564M-19a

Standard Specification for Hot-Rolled and Cold-Finished Age-Hardening Stainless Steel Bars and Shapes (Includes all amendments and changes 4/2/2021).

A693-22

Standard Specification for Precipitation-Hardening Stainless and Heat-Resisting Steel Plate, Sheet, and Strip

A705/A705M-23

Standard Specification for Age-Hardening Stainless Steel Forgings

F86-21

Standard Practice for Surface Preparation and Marking of Metallic Surgical Implants

F565-21

Standard Practice for Care and Handling of Orthopedic Implants and Instruments

F899-23

Standard Specification for Wrought Stainless Steels for Surgical Instruments

F983-86(2018)

Standard Practice for Permanent Marking of Orthopaedic Implant Components

F1264-16e1

Standard Specification and Test Methods for Intramedullary Fixation Devices (Includes all amendments and changes 1/24/2017).