ASTM F1800-97(2003) - 10.4.2003
 
1. Scope

1.1 This test method covers a procedure for the fatigue testing of metallic tibial trays used in knee joint replacements. This standard covers the procedures for the performance of fatigue tests on metallic tibial components using a cyclic, constant-amplitude force. It applies to tibial trays which cover both the medial and lateral plateaus of the tibia. This test method may require modifications to accommodate other tibial tray designs.

1.2 This test method is intended to provide useful, consistent, and reproducible information about the fatigue performance of metallic tibial trays with one unsupported condyle.

1.3 The values stated in SI units are regarded as the standard. The inch-pound units given in parentheses are for information only.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

 
2. Referenced Documents

E468-18

Standard Practice for Presentation of Constant Amplitude Fatigue Test Results for Metallic Materials

E1150-87(1993)

Definitions of Terms Relating to Fatigue (Withdrawn 1996)

E4-21

Standard Practices for Force Calibration and Verification of Testing Machines

E467-21

Standard Practice for Verification of Constant Amplitude Dynamic Forces in an Axial Fatigue Testing System