ASTM F2038-00(2011) - 1.12.2011
 
Significance and Use

4.1 This guide is intended to provide guidance for the specification and selection of silicone materials for medical device applications.

Silicone manufacturers supplying materials to the medical device industry should readily provide information regarding non-proprietary product formulation to their customers either directly, or through the US FDA master file program.

 
1. Scope

1.1 This guide is intended to educate potential users of silicone elastomers, gels, and foams relative to their formulation and use. It does not provide information relative to silicone powders, fluids, and other silicones. The information provided is offered to guide users in the selection of appropriate materials, after consideration of the chemical, physical, and toxicological properties of individual ingredients or by-products. This guide offers general information about silicone materials typically used for medical applications. Detail on the crosslinking and fabrication of silicone materials is found in Part II of this guide.

1.2 Fabrication and properties of elastomers is covered in the companion document, F604, Part II. This monograph addresses only components of uncured elastomers, gels, and foams.

1.3 Silicone biocompatibility issues can be addressed at several levels, but ultimately the device manufacturer must assess biological suitability relative to intended use.

1.4 Biological and physical properties tend to be more reproducible when materials are manufactured in accordance with accepted quality standards such as ANSI ISO 9001 and current FDA Quality System Regulations/Good Manufacturing Practice Regulations.

1.5 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Users are also advised to refer to Material Safety Data Sheets provided with uncured silicone components.

 
2. Referenced Documents

D1566-21a

Standard Terminology Relating to Rubber (Includes all amendments and changes 11/10/2021).

F813-20

Standard Practice for Direct Contact Cell Culture Evaluation of Materials for Medical Devices

AAMI TIRS-251

Microbiological Methods for Gamma Irradiation Sterilization of Medical Devices

21 CFR 211

Current Good Manufacturing Practice for Finished Pharmaceuticals (current revision)