ASTM F560-17 - 1.11.2017
 
1. Scope

1.1 This specification covers chemical, mechanical, and metallurgical requirements for unalloyed tantalum plate, sheet, strip, bar, and wire used in the manufacture of surgical implants.

1.2 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values of the two systems may result in non-conformance with this standard.

1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

 
2. Referenced Documents

E8-04

Standard Test Methods for Tension Testing of Metallic Materials

ISO 6892 Metallic Materials

Tensile Testing at Ambient Temperature

ASQ C1

Specifications of General Requirements for a Quality Program

F981-23

Standard Practice for Assessment of Muscle and Bone Tissue Responses to Long-Term Implantable Materials Used in Medical Devices

E29-22

Standard Practice for Using Significant Digits in Test Data to Determine Conformance with Specifications